MT4080A / MT4080D SPECIFICATIONS |
| |
Test Singal
Z Accuracy
D Accuracy
Q Accuracy
q Acceracy
Measuring Speed
General
Order Information
Standard/Optional Test Fixtures/Accessories, Part No. and Pictures |
|
| |
Measurement Functions |
|
| |
Measurement parameters:
Z : Absolute value of impedance
L : Inductance
C : Capacitance
DCR: Direct current resistance
ESR: Equivalent series resistance
D : Dissipation factor
Q : Quality factor
q : Phase angle
|
Combinations
| DCR |
Z |
Ls,Cs |
Lp,Cp |
| |
q |
D,Q,ESR |
D,Q |
Ranging: Auto-Range and Range-Hold
Relative: Mode
Cal: Open/Short calibration
RS232C:
Connector Type: IrDA
Transmission Singal: Tx and Rx
Flow Control: XON/XOFF |
|
| |
Test Singal
|
div align="right">
top
|
|
| |
Frequency: 100Hz, 120Hz, 1KHz, 10KHz, 100KHz(MT4080A only)±0.1% ( for both A&D )
Level: 1Vrms, 0.25Vrms, 0.05Vrms, 1Vdc(for DCR),±5%
Output Impedance: 100W, ±5%
Display Range
| |
Z
L
C
DCR
ESR
D
Q
q |
0.001W
0.000µH
0.000PF
0.000W
0.000W
0.000
0.000
-180.0° |
to
to
to
to
to
to
to
to |
9999MW
9999H
9999F
9999MW
9999W
9999
9999
180.0° |
|
|
| |
Z Accuracy |
|
|
| |
| DCR |
2% ± 1
u |
1% ± 1 |
0.5%
± 1 |
0.2%
± 1 |
0.5%
± 1 |
1% ± 1
u |
| 100Hz |
| 120Hz |
| 1KHz |
| 10KHz |
5% ± 1 |
2% ± 1 |
| 100KHz |
NA |
5% ± 1
u |
2% ± 1 |
0.4% ± 1 |
2% ± 1 |
5% ± 1
u |
Note :
1. The accuracy applies when the test level is set to 1Vrms.
2. Ae multiplies 1.25 when the test level is set to 250mVrms.
3. Ae multiplies 1.50 when the test level is set to 50mVrms.
4. When measuring L and C, multiply Ae by if the Dx>0.1.
u Ae is not specified if the test level is set to 50mV. |
|
| |
D Accuracy |
|
|
| |
D accuracy De is given as

Accuracy applies when Dx (measured D value) < 0.1
When DX > 0.1, multiply De by (1+Dx) |
|
| |
Q Accuracy |
|
| |
Q accuracy is given as

Where, Qx is the measure Q value, De is the relative D accuracy
Accuracy applies when Qx · De < 1 |
|
| |
q
Acceracy |
|
| |
q accuracy q is given as

|
|
| |
Measuring Speed |
|
| |
Fast : 4.5 meas. /sec
Slow : 2.5 meas. /sec |
|
| |
General |
|
|
| |
Temperature : 0°C to 40°C (Operating), 20°C
to 70°C (Storage)
Relative Humidity : Up to 85%
Battery Type : NiMh or Alkaline (2AA size)
Battery Charge : Constant current 150 mA approximately
Battery Operation Life : 2.5 Hours typical
AC Operation : 110V/220V AC, 60/50Hz with proper Adapter
Low Power Warning : under 2.2V
Dimensions : 6.9" x 3.4" x 1.9" (LxWxH) (174mm x 86mm x 48mm)
Weight : 470g |
|
| |
Order Information |
|
| |
MT4080A
100Hz to 100KHz Handheld LCR Meter
MT4080D
100Hz to 10KHz Handheld LCR Meter |
|
| |
Standard/Optional Test Fixtures/Accessories, Part No. and Pictures |
|
| |
Standard Accessories
Shorting bar
DC 6V power adapter
AA-Size Ni-Mh rechargeable battery Optional Test Fixtures/Accessories
TL08A SMD test probe
TL08B 4-wire test chips
TL08C 4-wire Kelvin chips
Carry case |
|
| |