Transmitter
TIMS (Transmission Impairment Measuring Set)
Impedance: 600W
Accuracy: < 5%
TX Frequency
Range: 404Hz, 1004Hz and 2804Hz
Accuracy: 1% @ 23°C ± 5°C
Temp. coefficient: 200ppm/°C
Distortion: 50dB down from fundamental
TX Level
Range: -13dBm or 0dBm
Accuracy: ± 0.3dB @ 23°C ± 5°C
Temp. coeficient: 200ppm/°C
Frequency Counter
Range: 0 to 20 KHz
Resolution: 10 Hz
Level: -40dBm to +20dBm
Accuracy: ± 0.5% reading ± 1 digit, -32dBm > level < 0dBm
Receiver
Impedance: 600W , 900W , or 1MW
Accuracy: < 1%
Other impedance available upon request
Rx Level
Range: (2)-50 to -12dBm or -12 to +8dBm
Resolution: 0.1dBm
Detector Type: True RMS
Accuracy: 0.3dB if:
1. 30Hz < Frequency < 3.4KHz
2. -20dBM < Level < +8dBm
Noise
Rx Range: -82 to -10dBm
Accurancy: ±2dBm, -72dBm to 10dBm
Detector Type: True RMS
Weighting Filter: C-MSG
DC Voltage
Range: 200V Full scale
Resolution: 0.1V
Accuracy: ± 0.5% reading ± 1 digit
Impedance: 1MW
AC voltage
Range: 300V Full Scale
Resolution: 1V
Accuracy: ± 0.5% reading ± 2 digit
Frequency response: 50Hz to 600Hz
Impedance: 1MW
Detector Type: True RMS
Resistance
Range: 2000W
Resolution: 1W
Accuracy: ± 0.5% reading ± 1 digit
Protection: 250VDC/250VAC max
Audible Continuity: < 30W
DC (loop) Current
Range: 200mA Full Scale
Resolution: 0.1mA
Accuracy: ± 0.1% reading ± 1 digit
Protection: 0.25A fuse
The specifications given assume an operating temperature of 73° F + 10°F.
Humidity up to 80%, and 1 year calibration cycle.
|